Olympus widefield for long term imaging


Purpose:
- Suitable for fluorescence, transmitted and phase imaging
- Suitable for live samples
- When no optical sectioning capability is required
- Large field-of-view observations
- Option for Microfluidics

Illumination / Imaging configuration: Widefield
Imaging contrasts: Brightfield, Phase contrast and Fluorescence
Microscope stand: Inverted Olympus IX81, motorized
Fluorescence excitation source WF: CoolLED pE-4000, LED light source 
Detection: 1 channel
Detection device:  Andor life 888 

Camera Andor Life 888
Technology EM-CCD
Chip size 1024x1024 pixels
Pixel size 13x13 µm
Dynamic range 16-bit
Max frame rate* 26 frames s-1
Quantum efficiency 95% @ 550 nm

*full chip, no binning

Acquisition software: µManager 2.0-gamma
Additional features:
XY motorized stage
- Hardware Z-drift compensation (Olympus ZDC2)
- Incubation box for temperature control, Air/CO2 mixing in option

Objectives:

Objective type Mag/NA Medium Contrast WB (mm) Z-Drift compensation system Pixel size (nm)
UPlanFLN

10x / 0.3

Air

Fluo, Ph

10

/

795

UPlanSApo 40x / 0.95

Air

Fluo

0.18

/

325

APON

60x / 1.4

Oil

Fluo

0.12

YES

216
UplanSApo

100x / 1.45

Oil

Fluo, Ph3

0.13

YES

130


Fluorescence filter cubes installed:

Fluorophore

Excitation filter

Dichroic filter cut-off

Emission filter

Filter cube reference

DAPI

387/11

409

447/60

DAPI-1160B

GFP

482/18

495

520/28

GFP-1828A

YFP

500/24

520

542/27

YFP-2427B

mCherry

562/40

593

593LP

mCherry-40LP-A